Annual Verification
Anritsu recommends an annual calibration and performance verification of the Site Master
and the OSL calibration components by local Anritsu service centers. Anritsu service cen
-
ters are listed in Table 1-2 on the following page.
The Site Master itself is self-calibrating, meaning that there are no field-adjustable compo
-
nents. However, the OSL calibration components are crucial to the integrity of the calibra
-
tion and therefore, must be verified periodically to ensure performance conformity. This is
especially important if the OSL calibration components have been accidentally dropped or
over-torqued.
1-7