Data convert function: Convert
The IQ sample data files in ASCII format programmed by
general signal generation software (such as M
ATLAB
) can also
be converted to waveform patterns for MG3700A.
This enhances the convenience of MG3700A for
research/development simulation use, since custom-made
waveform pattern files can be freely generated.
Convert screen
AWGN screen
File generation function: File Gen
AWGN (Additive White Gaussian Noise) waveform genera-
tion function
This function establishes the sampling rate and bandwidth,
allowing any AWGN waveform pattern to be created.
In addition, when the first combined waveform pattern (Wanted
Signal) is selected, the Wanted Signal bandwidth and sampling
rate are set automatically.
The resulting AWGN waveform pattern and an existing wave-
form pattern can be combined, which is useful for base-station
dynamic-range measurements.
The main set parameter
(1) Wanted Signal BW: Wanted Signal band-width
Setting range: 0.0010 to 120.0000 MHz
(2) AWGN BW (B)/Wanted Signal BW (A):
Magnification of AWGN to Wanted Signal
Setting range: 1.0, 1.5, 2.0, 2.5
(3) Sampling Rate:
Setting range: 0.0200 to 160.0000 MHz
It makes it to the same value as the Wanted Signal.
(4) AWGN BW (B): Bandwidth of AWGN
It calculates automatically from (1) and (2), and it has the
following limitation.
Limit range: 0.001 to 20.000 MHz and “Sampling rate/2” or
less than, 20.001 to 120.000 MHz and “Sampling rate” or
less than
EQ_MG3700A_E700 06.7.28 8:19 PM Page 23