Range 250 kHz to 3 GHz (standard), 250 kHz to 6 GHz (option)
Resolution 0.01 Hz
Frequency: 10 MHz, Aging rate: ±1 x 10
–8
/day, ±1 x 10
–7
/year, Temperature stability: ±2 x 10
–8
(0 to +50°C),
Start-up characteristics (at 23°C):
±5 x 10
–8
(After 5 min, compared to frequency after 24 h warm-up)
Internal reference oscillator With Rubidium Reference Oscillator Option
Frequency: 10 MHz, Aging rate: ±1 x 10
–10
/Month, Temperature stability: ±1 x 10
–9
(0 to +50°C),
Start-up characteristics (at 23°C):
±1 x 10
–9
(After 7 min, compared to frequency after 24 h warm-up)
External reference input
Frequency: 5 MHz/10 MHz (auto-switching), Operating range: ±1 ppm, Input level: 0.7 Vp-p/50
(AC coupled), Connector: BNC-J (rear panel, Ref Input)
Buffer output Frequency: 10 MHz, Output level: TTL (DC coupled), Connector: BNC-J (rear panel, Buffered Output)
Frequency
Response time from final command to ±0.1
ppm of set frequency on GPIB.
(When the set frequency is 1 GHz or less, response time from final command to ±100 Hz)
When Frequency changing speed = Normal:
40 ms (When going by 3 GHz)
15 ms (When the amount of frequency change is less than 1 GHz without going by 3 GHz)
20 ms (When the amount of frequency change is 1 GHz or more without going by 3 GHz)
Switching time When Frequency changing speed = Fast:
40 ms (When going by 3 GHz)
10 ms (When not going by 3 GHz)
With Mechanical Attenuator Option
Regardless of frequency changing speed.:
100 ms (When going by 3 GHz)
80 ms (When not going by 3 GHz)
Direct setup: The absolute value of frequency is set up and displayed.
Frequency setup and Display
Setup by CH: CH is assigned to frequency. Separate CH tables can be assigned to two or more
systems (groups). Group names and CH numbers are set and displayed. Furthermore, the
corresponding frequency is displayed simultaneously.
–140 to +13 dBm (At CW, accuracy range: –136 to +6 dBm)
With Mechanical Attenuator Option
Settable range
–140 to +19 dBm (At CW, accuracy range: –136 to +10 dBm)
:Please refer to “Vector modulation/In the case of vector modulation, the level error in comparison
with CW” for level accuracy at vector modulation.
Unit
Power: dBm,
Voltage: dBµV (terminate voltage display), dBµV (open voltage display)
Resolution 0.01 dB (dBm, dBµV)
at CW and 23°C ±5°C:
Output level
Accuracy
“Upper frequency 6 GHz option” is required for “3 GHz < f 6 GHz.
With Mechanical Attenuator Option
“Upper frequency 6 GHz option” is required for “3 GHz < f 6 GHz.
MG3700A mainframe
1
Specifications
Level (p) [dBm]
Frequency (f) [Hz]
250 k f < 25 M 25 M f 3 G 3 G < f 6 G
+3 < p +6 ±0.5 dB
–1 < p +3 ±0.5 dB ±0.8 dB
–120 p –1 ±0.5 dB typical ±0.5 dB ±0.8 dB
–127 p <–120 ±0.7 dB ±2.5 dB typical
–136 p <–127 ±1.5 dB typical
Level (p) [dBm]
Frequency (f) [Hz]
250 k f < 25 M 25 M f 3 G 3 G < f 6 G
+7 < p +10 ±0.5 dB typical ±0.5 dB
–100 p +7 ±0.5 dB typical ±0.5 dB ±0.8 dB
–120 p <–100 ±0.5 dB typical ±0.5 dB ±1.0 dB
–127 p <–120 ±0.7 dB ±2.5 dB typical
–136 p <–127 ±1.5 dB typical